Tracking accuracy evaluation of electromagnetic sensor-based colonoscope tracking method

Guardat en:
Dades bibliogràfiques
Publicat a:arXiv.org (Jan 14, 2022), p. n/a
Autor principal: Oda, Masahiro
Altres autors: Kondo, Hiroaki, Kitasaka, Takayuki, Furukawa, Kazuhiro, Miyahara, Ryoji, Hirooka, Yoshiki, Goto, Hidemi, Navab, Nassir, Mori, Kensaku
Publicat:
Cornell University Library, arXiv.org
Matèries:
Accés en línia:Citation/Abstract
Full text outside of ProQuest
Etiquetes: Afegir etiqueta
Sense etiquetes, Sigues el primer a etiquetar aquest registre!
Descripció
Resum:This paper reports a detailed evaluation results of a colonoscope tracking method. A colonoscope tracking method utilizing electromagnetic sensors and a CT volume has been proposed. Tracking accuracy of this method was evaluated by using a colon phantom. In the previously proposed paper, tracking errors were measured only at six points on the colon phantom for the accuracy evaluation. The point number is not enough to evaluate relationships between the tracking errors and positions in the colon. In this paper, we evaluated the colonoscope tracking method based on more detailed measurement results of the tracking errors. We measured tracking errors at 52 points on the colon phantom and visualized magnitudes of the tracking errors. From our experiments, tracking errors in the ascending and descending colons were enough small to perform colonoscope navigations. However, tracking errors in the transverse and descending colons were large due to colon deformations.
ISSN:2331-8422
DOI:10.1007/978-3-319-29965-5_10
Font:Engineering Database